| Instrument | Location | Range & Accuracy |
Air Sensor | Pre-column | Ultrasonic (0–10 mm) |
Liquid Sensor | Air trip | Green/red |
Pressure transmitter | Post Pump, Pre & Post-Column | -0.1 ~ 4 bar |
Conductivity Sensor | Pre & Post-Column | 0.1 - 200 mS/cm (± 0.25 mS/cm) 0-100μS/cm(±3μS/cm) |
Temperature Sensor | Integrated on Cond Sensor | 5 -40°℃,(±1°℃) |
pH Sensor | Pre-column (Optional), Post-Column | 3~10,±0.1pH |
| UV Sensor | Post-Column | 0-3 AU @ 280 nm, ±0.5% 0 - 3 AU @ 254 nm, ± 0.5% |
| Flowmeter | Pre-column, Post-column (Optional) | 3% of Range |
Dual-pump gradient configuration
Disposable sensors requiring no calibration
Fully disposable fluid path; material certificates for fluid path kits are available upon request
Fluid path kits with varying tubing diameters can be selected based on flow rate requirements
Disposable fluid path design facilitates rapid installation and removal, eliminating the challenges associated with cleaning validation
Design and documentation compliant with cGMP requirements
Materials compliant with USP Class VI standards, accompanied by comprehensive validation documentation
Intuitive process step editor and automation platform
System software complies with the requirements of GAMP 5 and FDA 21 CFR Part 11 regulations
Users can configure operating parameters—including feed volume, flow rate, pressure, etc.—to suit their specific process requirements
Flexible and powerful recipe editing capabilities, supporting both text-based editing and text-based previewing of recipes
Software features such as user management, audit trails, and data export functions ensure data integrity
Multiple collection modes available for selection
Real-time display of operating parameters and process curves
Data analysis module featuring automated column efficiency calculation